Rotronic

Features

  • Developed for the use with Rotronic PT100 probes
  • Suitable with all Rotronic Binder 4 pole connector in combination with handheld (HP32) or transmitter (HF5, PF4, PF5)
  • Extended temperature measurement range from -100 up to 400 °C
Storage Conditions -40…85 °C
Weight 19.1 g
Current consumption Approx. 5 mA (adjusted at 3.3 VDC)
Power supply 3.3…5 VDC
Resolution analog output 15 bit
Load per analog output V-signal: ≥1 kΩ/V
Digital interface options UART (19200 baud fixed)
Analog outputs Humidity (disabled)
Temperature (default)
Analog output scaling Humidity (disabled) Humidity (disabled)
  • Temperature (-40…60 = 0…1V)
  • Temperature freely settable with HW5
Protocols RoASCII (default)
MODBUS (setting with HW5)
IP protection class IP65
Measurement uncertainty 0.1°C
Timing 1st measurement after 1.5 s
Measurement interval of 1.1 s

Filtering

To share